CVE-2020-12045The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded creden…
Updated under an hour ago ()
Does this affect you?
It affects baxter sigma_spectrum_infusion_system_firmware. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 1.68%, higher than 77% of scored CVEs.
Do this next
Free account, 1 product
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.
- CVSS
- 9.8 CRITICAL (v3.1, NVD)CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- Weakness (CWE)
- CWE-259
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 1.68% (percentile 77)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Aug 4, 2024
- First seen by InfoSec Signals
- Oct 12, 2026
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.