CVE-2020-25150B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
Updated under an hour ago ()
Does this affect you?
It affects B. Braun Melsungen AG SpaceCom and 4 other products. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 1.53%, higher than 75% of scored CVEs.
Do this next
Free account, 1 product
A relative path traversal attack in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with service user privileges to upload arbitrary files. By uploading a specially crafted tar file an attacker can execute arbitrary commands.
- CVSS
- 7.6 HIGH (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L
- Weakness (CWE)
- CWE-23
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 1.53% (percentile 75)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Apr 16, 2025
- First seen by InfoSec Signals
- Oct 12, 2026
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.