CVE-2021-23236Fresenius Kabi Agilia Connect Infusion System uncontrolled resource consumption
Updated under an hour ago ()
Does this affect you?
It affects Fresenius Kabi Agilia Link+ and 6 other products. Add what you run to find out whether that includes you.
EPSS puts the chance of exploitation in the next 30 days at 1.07%, higher than 65% of scored CVEs.
Do this next
Free account, 1 product
Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system.
- CVSS
- 7.5 HIGH (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
- Weakness (CWE)
- CWE-400
- Known exploited
- Not in CISA's KEV catalog
- EPSS (exploitation in the next 30 days)
- 1.07% (percentile 65)
- Assigned by
- icscert
- NVD analysis
- Modified
- Record last updated
- Apr 16, 2025
- First seen by InfoSec Signals
- Oct 12, 2026
The CVE record on cve.org NVD's entry
The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.