Back to CVE records
Critical 9.8HealthcarePublished Jun 1, 2022

CVE-2022-29875A vulnerability has been identified in Biograph Horizon PET/CT Systems (All VJ30 versions < VJ30C-UD01), MAGNETOM Family (NUMARIS X: VA12M, VA12S, VA10B, VA20A, VA30A, VA31A), MAMMOMAT Revelation (Al…

Updated under an hour ago ()

Does this affect you?

It affects Siemens Biograph Horizon PET/CT Systems and 39 other products. Add what you run to find out whether that includes you.

EPSS puts the chance of exploitation in the next 30 days at 1.87%, higher than 79% of scored CVEs.

A vulnerability has been identified in Biograph Horizon PET/CT Systems (All VJ30 versions < VJ30C-UD01), MAGNETOM Family (NUMARIS X: VA12M, VA12S, VA10B, VA20A, VA30A, VA31A), MAMMOMAT Revelation (All VC20 versions < VC20D), NAEOTOM Alpha (All VA40 versions < VA40 SP2), SOMATOM X.cite (All versions < VA30 SP5 or VA40 SP2), SOMATOM X.creed (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.All (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Now (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Open Pro (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Sim (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Top (All versions < VA30 SP5 or VA40 SP2), SOMATOM go.Up (All versions < VA30 SP5 or VA40 SP2), Symbia E/S (All VB22 versions < VB22A-UD03), Symbia Evo (All VB22 versions < VB22A-UD03), Symbia Intevo (All VB22 versions < VB22A-UD03), Symbia T (All VB22 versions < VB22A-UD03), Symbia.net (All VB22 versions < VB22A-UD03), syngo.via VB10 (All versions), syngo.via VB20 (All versions), syngo.via VB30 (All versions), syngo.via VB40 (All versions < VB40B HF06), syngo.via VB50 (All versions), syngo.via VB60 (All versions < VB60B HF02). The application deserialises untrusted data without sufficient validations that could result in an arbitrary deserialization. This could allow an unauthenticated attacker to execute code in the affected system if ports 32912/tcp or 32914/tcp are reachable.

CVSS
9.8 CRITICAL (v3.1, NVD)CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness (CWE)
CWE-502
Known exploited
Not in CISA's KEV catalog
EPSS (exploitation in the next 30 days)
1.87% (percentile 79)
Assigned by
siemens
NVD analysis
Modified
Record last updated
Aug 3, 2024
First seen by InfoSec Signals
Oct 12, 2026

Affected products

Affected products
Siemens Biograph Horizon PET/CT Systems
siemens:biograph_horizon_pet_ct_systems (from the name)
All VJ30 versions < VJ30C-UD01
CNA record
Siemens MAGNETOM Family
siemens:magnetom_family (from the name)
NUMARIS X: VA12M, VA12S, VA10B, VA20A, VA30A, VA31A
CNA record
Siemens MAMMOMAT Revelation
siemens:mammomat_revelation (from the name)
All VC20 versions < VC20D
CNA record
Siemens NAEOTOM Alpha
siemens:naeotom_alpha (from the name)
All VA40 versions < VA40 SP2
CNA record
Siemens SOMATOM X.cite
siemens:somatom_x.cite (from the name)
All versions < VA30 SP5 or VA40 SP2
CNA record
Siemens SOMATOM X.creed
siemens:somatom_x.creed (from the name)
All versions < VA30 SP5 or VA40 SP2
CNA record
Siemens SOMATOM go.All
siemens:somatom_go.all (from the name)
All versions < VA30 SP5 or VA40 SP2
CNA record
Siemens SOMATOM go.Now
siemens:somatom_go.now (from the name)
All versions < VA30 SP5 or VA40 SP2
CNA record
Siemens SOMATOM go.Open Pro
siemens:somatom_go.open_pro (from the name)
All versions < VA30 SP5 or VA40 SP2
CNA record
Siemens SOMATOM go.Sim
siemens:somatom_go.sim (from the name)
All versions < VA30 SP5 or VA40 SP2
CNA record

References

References

The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.