Back to CVE records
High 7.2HealthcarePublished Dec 17, 2025

CVE-2025-14097Remote Code Execution Vulnerability in Radiometer Products

Updated under an hour ago ()

Does this affect you?

It affects Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers and 2 other products. Add what you run to find out whether that includes you.

EPSS puts the chance of exploitation in the next 30 days at 0.46%, higher than 40% of scored CVEs.

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

CVSS
7.2 HIGH (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
Weakness (CWE)
CWE-287
Known exploited
Not in CISA's KEV catalog
EPSS (exploitation in the next 30 days)
0.46% (percentile 40)
Assigned by
Radiometer
NVD analysis
Deferred
Record last updated
Dec 17, 2025
First seen by InfoSec Signals
Oct 9, 2026

Affected products

Affected products
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers
radiometer_medical_aps:abl90_flex_and_abl90_flex_plus_analyzers (from the name)
Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems; Application software versions >= 3.5MR11 with Windows 10 operating system not affected
CNA record
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers
radiometer_medical_aps:abl800_basic_and_abl800_flex_analyzers (from the name)
Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems; Application software versions >= 6.20MR2 with Windows 10 operating system not affected
CNA record
Radiometer Medical Aps AQT90 FLEX Analyzers
radiometer_medical_aps:aqt90_flex_analyzers (from the name)
Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems; Application software versions >= 8.13 MR2 with Windows 10 operating system not affected
CNA record

References

References

The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.

Your cookie choices
We use essential cookies to run this site, and, only with your consent, an advertising cookie from Google to measure whether our ads lead to sign-ups and subscriptions. See our for details.