Back to CVE records
High 8.4Published Aug 5, 2026

CVE-2026-17583Thermo Fisher Applied Biosystems Genetic Analyzers Missing Support for Integrity Check

Updated under an hour ago ()

Does this affect you?

It affects Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software and 15 other products. Add what you run to find out whether that includes you.

No exploitation reported and no EPSS estimate yet.

The affected Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.

CVSS
8.4 HIGH (v3.1, CNA)CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness (CWE)
CWE-353
Known exploited
Not in CISA's KEV catalog
EPSS (exploitation in the next 30 days)
Not loaded for this CVE
Assigned by
icscert
NVD analysis
Analyzed
Record last updated
Aug 6, 2026
First seen by InfoSec Signals
Oct 9, 2026

Affected products

Affected products
Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software
thermo_fisher:applied_biosystems_3500_3500xl_series_data_collection_software (from the name)
through 4.0.2
CNA record
Thermo Fisher Applied Biosystems 3730/3730xL Series Data Collection Software
thermo_fisher:applied_biosystems_3730_3730xl_series_data_collection_software (from the name)
through 5.0.2
CNA record
Thermo Fisher Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software
thermo_fisher:applied_biosystems_seqstudio_genetic_analyzer_data_collection_software (from the name)
through 1.2.5
CNA record
Thermo Fisher Applied Biosystems SeqStudio Flex Series Instrument Software
thermo_fisher:applied_biosystems_seqstudio_flex_series_instrument_software (from the name)
through 1.2.0
CNA record
Thermo Fisher Applied Biosystems GeneMapper ID-X Software
thermo_fisher:applied_biosystems_genemapper_id_x_software (from the name)
through 1.7.3
CNA record
Thermo Fisher Applied Biosystems 3130 Series Data Collection Software
thermo_fisher:applied_biosystems_3130_series_data_collection_software (from the name)
through 4.1
CNA record
Thermo Fisher ABI PRISM 3100/3100-Avant Data Collection Software
thermo_fisher:abi_prism_3100_3100_avant_data_collection_software (from the name)
through 2.0
CNA record
Thermo Fisher ABI PRISM 310 Data Collection Software
thermo_fisher:abi_prism_310_data_collection_software (from the name)
through 3.1
CNA record
thermofisher abi_prism_310_data_collection_software
thermofisher:abi_prism_310_data_collection_software
through 3.1
NVD
thermofisher abi_prism_3100/3100-avant_data_collection_software
thermofisher:abi_prism_3100/3100-avant_data_collection_software
through 2.0
NVD

References

References

The CVE record is authoritative. Products marked “from the name” were identified from the CNA's free-text names, not a CPE, and are matched less certainly. This product uses the NVD API but is not endorsed or certified by the NVD.