Boston Scientific Corporation: System, Image Processing, Radiological
Updated under an hour ago ()
Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
- Tags
- FDA recall, medical device, class 2, Radiology, Device Design, Open, Classified
- Related exam domains
- CISSP 3: Security Architecture and Engineering; CISSP 8: Software Development Security
- Source record id
- Z-2771-2026
- First seen by InfoSec Signals
- Sep 23, 2026
Source: openFDA device recalls, U.S. Food and Drug Administration about this source more from it
Exam domain labels come from a keyword heuristic and are study hints, not an official mapping. The summary is the publisher's own text, shortened; the linked record is authoritative.