Beckman Coulter, Inc.: Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cytometric Analysis

Updated under an hour ago ()

Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.

Tags
FDA recall, medical device, class 1, Clinical Chemistry, Software design, Open, Classified
Related exam domains
CISSP 3: Security Architecture and Engineering; CISSP 8: Software Development Security
Source record id
Z-0412-2026
First seen by InfoSec Signals
Sep 23, 2026

Source: openFDA device recalls, U.S. Food and Drug Administration about this source more from it