Beckman Coulter, Inc.: Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cytometric Analysis
Updated under an hour ago ()
Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.