Wipro GE Healthcare Private Ltd.: Display, Cathode-Ray Tube, Medical

Updated 2 hours ago ()

Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

Tags
FDA recall, medical device, class 2, Cardiovascular, Software design, Open, Classified