FDA device recallVulnerabilitiesPublished Oct 8, 2025

Canon Medical System, USA, INC.: Interventional Fluoroscopic X-Ray System

Updated under an hour ago ()

It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.

Tags
FDA recall, medical device, class 2, Radiology, Radiation Control for Health and Safety Act, Open, Classified