Back to device advisories
FDA device recallVulnerabilitiesPublished Sep 26, 2025

Zap Surgical Systems: Accelerator, Linear, Medical

Updated under an hour ago ()

If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.

Tags
FDA recall, medical device, class 2, Radiology, Software change control, Open, Classified
Related exam domains
CISSP 3: Security Architecture and Engineering; CISSP 8: Software Development Security; CISM 3: Information Security Program
Source record id
Z-2639-2025
First seen by InfoSec Signals
Sep 23, 2026

Source: openFDA device recalls, U.S. Food and Drug Administration about this source more from it

Exam domain labels come from a keyword heuristic and are study hints, not an official mapping. The summary is the publisher's own text, shortened; the linked record is authoritative.

Your cookie choices
We use essential cookies to run this site, and, only with your consent, an advertising cookie from Google to measure whether our ads lead to sign-ups and subscriptions. See our for details.