Medtronic MiniMed, Inc.: Continuous Glucose Monitor Retrospective Data Analysis Software

Updated under an hour ago ()

CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).

Tags
FDA recall, medical device, class 1, Clinical Chemistry, Software design, Open, Classified
Related exam domains
CISSP 3: Security Architecture and Engineering; CISSP 8: Software Development Security; CISA 3: Information Systems Acquisition, Development and Implementation; CISA 5: Protection of Information Assets
Source record id
Z-0203-2027
First seen by InfoSec Signals
Oct 23, 2026

Source: openFDA device recalls, U.S. Food and Drug Administration about this source more from it

Exam domain labels come from a keyword heuristic and are study hints, not an official mapping. The summary is the publisher's own text, shortened; the linked record is authoritative.

Your cookie choices
We use essential cookies to run this site, and, only with your consent, an advertising cookie from Google to measure whether our ads lead to sign-ups and subscriptions. See our for details.